Exécution de notre outil de synthèse sur un nouvel article

Vérifiez l’article ci-dessous, puis lancez sa synthèse.


A 0.5V PPG-based Heart Rate and Variability Detection System

Wala Saadeh (Lahore University of Management Sciences), Shah Zaib Aslam (Lahore University of Management Sciences), Aminah Hina (Lahore University of Management Sciences), F. Fard Ali Asghar (Lahore University of Management Sciences)

IEEE Biomedical Circuits and Systems Conference (BioCAS)
Licence : non indiquée (tous droits réservés)

Résumé : Heart Rate Variability (HRV) is a diagnostic and predictive tool for multiple health conditions. A Photoplethysmographic (PPG) based HRV detection system with a low-power readout together with an embedded feature extraction processor to measure heart rate (HR) and HRV is presented. The proposed system proposes a low power LED driver with novel clock bootstrapping switching. The LED current is digitally controlled for adjusting LED brightness using Pulse Density Modulation (PDM) with >90% efficiency. It also introduces a highly compact low power/low noise transimpedance amplifier for an input base current range (0.5nA-4.5μA). The proposed system eliminates the motion artifacts without the need for an accelerometer. The complete system is simulated using 65nm CMOS technology with 0.5V supply. It consumes 3 μW/1.1mW of power for the HRV detection and LED driver, respectively.

Publié le 01 Oct. 2018